תיאור
Fully Equipped
This vacuum-tight STA 449 F5 Jupiter® is hand built specifically for your application. Each instrument comes pre-configured as you need it, with all of the hardware and software features that are required by high-temperature applications in the fields of ceramics, metals, inorganics, building materials, etc.
Best Cost-Performance-Ratio
The STA 449 F5 Jupiter® has high resolution, a high range, low drift balance and sensitive DSC capability that enables it to excel in all typical applications across a broad range of temperatures.
Unique Combination – True TG-DSC and TGA for High-Sample Volumes
True TGA and DSC measurements can be performed with high precision and reproducibility at sample temperatures from ambient to 1600°C. TGA measurements are also possible, even on large or heavy samples.
Top-loading – The Proven Design for Thermobalances
The STA 449 F5 Jupiter® is a top-loading system with a balance design that has been preferred for years in laboratories. The reasons are simple: These systems combine ideal performance with easy handling.
- Temperature range:
RT to 1600°C (sample temperature) - Heating rate:
0.001 to 50 K/min - Vacuum-tight:
10-2 mbar - Temperature resolution:
0.001 K - Balance resolution:
0.1 μg (over the entire weighing range) - Balance drift:
< 5 μg/hour - Maximum sample load:
35000 mg (incl. crucible), corresponds to TGA measuring range - Sample volume:
Up to 5 cm3 (for TGA crucibles) - Enthalpy accuracy:
1% (indium)